Instruction ManualPart Numbers: B7150 and B7350ScanX-12 DVMP/N B7350ScanX DVMP/N B7150
10COMPUTER SYSTEM REQUIREMENTSIMPORTANT: To operate the ScanX, it must be connected to a compliant Computer System, and the computer must be loaded
11COMPUTER SYSTEM REQUIREMENTSCONTROLS AND INDICATORSALLPROIMAGINGREADY ERASERFLASHINGOFFALLPROIMAGINGREADY ERASERFLASHINGOFFREADY SwitchFLASHING OFF
12TECHNICAL DATAElectrical Requirements:Supply Voltage: 100 to 240VAC +/- 10%, 50/60 Hz Supply Current: 1.2 A MaximumLine Cord: North American sty
13The ScanX may be located almost anywhere in the office. Follow these guidelines for optimum performance: Lighting conditions: Set up the scanner i
14SYSTEM SETUPNote: Authorized Imaging Software, such as Visix or VetRay, supplied by the dealer or other company, must be installed on the computer
15Figure 1. Typical ScanX InstallationSYSTEM SETUPPower SupplyComputer Connector CableMains OutletLine CordNote:The Line Cord is the Mains disconnect
16Printed sideTube or Sensitive sideFigure 2. Intraoral PSP ConfigurationPrior to performing the Intraoral and Extraoral imaging procedures provi
17PLATE CARE & PREPARATIONIMPORTANT: PSPsmustalwaysbeerasedpriortouse.Note: UsePSPswithin24hoursoflasterasure.Repeaterasingproc
18Figure 4aInsert the PSP.Figure 4bPeel off the adhesive strip.Figure 4cSeal the Barrier Envelope.Preparing Intraoral Plates for Patient UseInsert the
19Note: 1. The orientation letter “a”, printed on the PSP, may be used for reference as you would use the dot on an intraoral X-ray film. In additi
TABLE OF CONTENTS2FOREWORDALLPRO Imaging has prepared this document as a guide to the proper use of two ScanX® Digital Imaging Systems with In-Line Er
204. Use the user-supplied authorized Imaging Software to activate the Scanner and to select the desired image type and resolution. 5. Verify t
21INTRAORAL IMAGING PROCEDURESScanning and Erasing Plates (continued)Note: Up to four PSPs can be processed simultaneously. One PSP can be inserted i
22EXTRAORAL IMAGING PROCEDURESFigure 11.Feeding an Extra Oral Plate into the ScanX1. Orient the cassette so that the Tube side is fac-ing down and th
23EXTRAORAL IMAGING PROCEDURESScanning and Erasing Plates (continued)5. Observe that a red glow eminates from the scanner exit slot.6. Repeat steps
24MAINTENANCE Maintenance ProceduresThe ScanX is designed for many years of trouble-free operation. Maintenance as described herein is minimal.IMPORTA
25Like all precision products, the ScanX requires a certain amount of care on a regularly scheduled basis. A well-organized maintenance p
26TROUBLESHOOTINGNote: There should be an image on the plate only if the in-line eraser mode was disabled during scanning.Trouble Possible Cause Corre
27TROUBLESHOOTINGTrouble Possible Cause Corrective Action6. Image is too dark.• PSP has been overexposed • Use software to adjust bright-ness. I
28The following lists the ordering number and description for accessory components available to maintain the ScanX to meet your professional needs. Co
29VISIX IMAGING SOFTWAREConsider VISIX, our comprehensive digital imaging software application, providing seamless integration with our ScanX P
3CONGRATULATIONSCongratulations on your purchase of the ScanX-12 DVM or ScanX DVM Digital Imaging System with In-Line Erase another innovati
© Copyright ALLPRO Imaging, 2006 • P/N B7157, Rev. D - December 2010ALLPRO Imaging is a leading manufacturer of imaging systems for the medical, veter
SAFETY NOTICE4This equipment has been designed to minimize exposure of personnel to haz-ards. While the ScanX is designed for safe operation, cer
5SAFETY NOTICEUse of Accessory Equipment.The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipm
6Knowledge of Warnings and Cautions.Users must exercise every precaution to ensure personnel safety, and be familiar with the warnings and cautions pr
7General Notes. All instructions in this manual form an integral part of the unit. They must be kept close to the unit and in readiness whe
8IMPORTANT INFORMATIONCorrect Usage Operation of the ScanX may only be carried out by suitably qualified personnel. The ScanX is only to be used i
9UNPACKING AND INSPECTIONUnpackingThe ScanX may be shipped in a very special shipping case covered by a double-wall corrugated over-packaging. A depos